NOA 280i Specifications

Operating Specifications

Operating Specifications
Sensitivity Gas: < 1 ppb

Liquid: ~1 picomole

Range Gas: <1–500,000 ppb

Liquid: Nanomolar to millimolar

Response Time Electronics: 67 milliseconds to 90% full scale

Lag Time: 1 second

Repeatability Gas: ± 5%

Liquid: ± 5–10%

Sample Size Gas: 10–300 mL/min

Liquid: 0.001–5 mL

Display Back-lit LCD screen

ppb/ppm or mV

Outputs Analog: 0–1V, 0–10 V

Digital: RS-232 (9600-38.4K baud)

Printer: Parallel port

Data Sampling Rate 0.002–32 samples/second
Power Requirements 120 V, 60 Hz (6A)

100 V, 50 or 60 Hz (7A)

230 V, 50 Hz (3A)

NOA Height: 16 in (41 cm)

Width: 6.2 in (16 cm)

Length: 20 in (51 cm)

Weight: 35 lbs (16 kg)

Vacuum Pump with

Installed Trap

Height: 14.5 in (37 cm)

Width: 7.5 in (19 cm)

Length: 19 in (48 cm)

Weight: 47 lbs (21.5 kg)

Certifications Declaration of conformity
Operating Environment
Ambient Temperature 32° to 86°F (0° to 30°C)
Relative Humidity 0–90%
Fuse Requirements
Main Fuse 100 VAC model: T, 10A, 250V

120 VAC model: T, 5A, 250V

230 VAC model: T, 5A, 250V

Ozone Fuse 100 VAC model: T, 200mA, 250V

120 VAC model: T, 200mA, 250V

230 VAC model: T, 100mA, 250V